An excellent technical article about calibration standard definitions by William Highton (Chemandy Electronics) can be found here.
The main reason for matching the calibration reference plane and the mating plane of the VNA test port connector is to allow industrial VNAs to measure whole assemblies and devices, including their connectors. In the N2PK VNA (at least with the original software) the reference plane is assumed to be where the physical OSL calibration takes place. Instead of adding negative offset to move the reference plane back to the mating surface of the VNA's test port connector, it is accepted that the plane is right at the rear surface of the calibration standard connector (the location of the actual Open-Short-Load). The error introduced by the electrical length of the pin inside the connector is not that great at lower frequencies - for instance, assuming teflon dielectric (VF ~70%) in a standard SMA connector used for a home-brewed calibration standard, an offset of 5 mm will yield about 0.5 degree phase error @60 MHz. Commercial calibration standards use air dielectric, but this puts greater mechanical requirements during the fabrication process. By calibrating the VNA so the reference plane is on the back of the cal connector and fully exposed, it is more convenient to measure a single component but it means that the DUT test fixture should be made of (or at least include) the same type connector as the one used for the home-brewed calibration standards in order to maximize the accuracy. The setup can be as simple as the component (DUT) is just soldered on the back of the "test fixture" connector, while measures are taken to minimize external error (minimal lead length). With SMDs, this approach works well as the device is practically located directly at the "reference plane" when the test fixture connector is from the same batch connectors used to make the OSL standards. Another method for calibration is to use the test fixture connector itself for calibration - for OPEN the DUT is simply omitted, SHORT and LOAD are created on the spot during each calibration step and finally the DUT is soldered (Still, stray C for the OPEN must be known)

2 comments:
You state:
"Using a calibration standard which mates with such adapter will move the calibration reference plane at (or close to) the DUT, "absorbing" the adapter in the calibration process.
BUT the calibration kit used should be that of the DUT, not the VNA. So you should put the adapter on the cable, then calibrate after the adapter using a calibration kit for whatever connector your DUT has.
This is exactly right! There is no need for "BUT" :-) as there is no argument here!
The only reason I worded this as such is because many DIY VNA owners focus at making a calibration kit matching the VNA's native connector. What I was trying to say, is that one should have sets of Cal Standards, for each possible DUT connector type to be used. Then, any needed cable extensions and adapters will be taken out of the measurment process as the reference plane is moved by the location of the cal standard.
Thanks for summrizing it a more clear way.
Post a Comment